1942 4-12UP LEAK TESTER

1942 4-12UP LEAK TESTERDownload .pdf

The 1942 4-12UP Leak Tester is used for the simultaneous final inspection of up to twelve balloon catheters. The product is subjected to a specific pressure or vacuum and held for a defined time. The corresponding LED flashes when the pressure drops below the specification.

The process parameters can be entered and managed via the user interface. The process data are displayed directly and can be visualised graphically in the test report. The process can consist of up to five parameter steps, which start automatically or manually.

The balloons can be deflated with a vacuum pump after the test.

Technical Specifications
test connections: Luer male
pressure range: 0–48 bar (700 psi)
pressure control accuracy: +/– 0.25 bar (+/– 4 psi)
measurement accuracy: 0.01 bar (0.15 psi)
vacuum: diaphragm pump 230 mbar (3.3 psi) absolute
medium: air or N2
process parameters: set pressure, pressure difference, pressure ramp, holding time and measuring time
dimensions (L x W x H): 800 x 404 x 400mm
weight: 25kg
power: 240W
voltage: 110–240 VAC, 50 –60Hz
versions: 4UP, 8UP, 12UP

Details:

Luer Connectors
Interfaces
BW-TEC HMI